Instrument Operation Flashcards

7 cards from real SCYM practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Instrument Operation flashcards as text
  1. What is the primary purpose of the sheath fluid in a flow cytometer?

    Answer: To hydrodynamically focus cells into a single-file stream

    Sheath fluid creates laminar flow that hydrodynamically focuses the sample stream so cells pass through the laser interrogation point one at a time.

  2. Which parameter is used to distinguish doublets from single cells during flow cytometric analysis?

    Answer: FSC height vs FSC area (or width)

    Doublet discrimination is performed by comparing FSC-H (height) vs FSC-A (area); doublets produce disproportionately higher area relative to height.

  3. When performing daily QC on a flow cytometer using fluorescent beads, a sudden rightward shift in bead peak CV is most likely caused by:

    Answer: Laser power fluctuation

    Laser power instability causes variable fluorescence excitation, broadening the bead peak and increasing CV.

  4. In a spectral flow cytometer, unmixing algorithms require which reference material?

    Answer: Single-color controls (single-stained references) for each fluorochrome

    Spectral unmixing requires single-color reference spectra for each fluorochrome to mathematically separate overlapping emission signals across detectors.

  5. The time delay setting in a cell sorter is used to:

    Answer: Synchronize the deflection charge with the droplet containing the target cell

    The time delay (sort delay) accounts for the transit time from the laser interrogation point to the droplet break-off point so the correct droplet is charged.

  6. Which action should be taken first if the sample flow rate on a benchtop cytometer drops unexpectedly mid-run?

    Answer: Check for a clog in the sample line and perform a backflush if needed

    A sudden flow rate drop typically indicates a partial clog; checking and performing a backflush resolves most sample-line obstructions before escalating.

  7. What does a low coefficient of variation (CV) for a fluorescent bead peak indicate during instrument QC?

    Answer: The instrument is performing consistently with tight measurement precision

    A low CV on bead peaks indicates that the instrument is producing highly reproducible fluorescence measurements, reflecting good optical and electronic performance.