โ† All RAC Flashcard Decks

Post-Market Surveillance Flashcards

7 cards from real RAC practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Post-Market Surveillance flashcards as text
  1. Under EU MDR 2017/745, what is a 'Field Safety Corrective Action' (FSCA)?

    Answer: An action taken to reduce risk of death or serious deterioration in health associated with a device already on the market

    An FSCA is any action taken to reduce a risk of death or serious deterioration related to a device in use, and must be reported to competent authorities.

  2. Which metric is MOST useful for quantifying the real-world safety signal in post-market surveillance data?

    Answer: Complaint rate per units distributed

    Expressing complaints as a rate per units distributed normalizes the data, allowing meaningful comparison across time periods and product versions.

  3. Under 21 CFR Part 803, a device manufacturer must report a device-related death within how many calendar days of becoming aware?

    Answer: 30 days

    Manufacturers must submit an MDR for deaths and serious injuries within 30 calendar days of becoming aware, or 5 days if an immediate hazard is identified.

  4. Which of the following is a proactive PMS data source, as distinguished from a reactive one?

    Answer: Literature surveillance reviewing published clinical studies

    Literature surveillance is proactive because it monitors published evidence regardless of whether an adverse event has already occurred with the manufacturer's specific device.

  5. A company receives a complaint that a sterile device's packaging was found breached upon receipt. What PMS action is MOST critical to initiate first?

    Answer: Assess the complaint for MDR reportability and initiate a CAPA investigation

    Packaging integrity failures affecting sterility are potential MDRs and require formal CAPA investigation to determine root cause and prevent recurrence.

  6. What distinguishes a Post-Market Surveillance Report (PMSR) from a Periodic Safety Update Report (PSUR) under EU MDR?

    Answer: PMSRs are for Class I devices only; PSURs are required for Class IIa, IIb, and III

    Under EU MDR Articles 85 and 86, Class I devices need a PMSR while Class IIa, IIb, and III devices require the more comprehensive PSUR.

  7. In the context of PMS complaint handling, what does 'MDR decisioning' require the manufacturer to document?

    Answer: The rationale for why a complaint does or does not meet MDR reporting criteria

    FDA expects written documentation of the analysis and justification used to determine whether each complaint meets MDR reportability thresholds.