Testing and Quality Assurance Flashcards

7 cards from real ECHOCARDIOGRAM practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Testing and Quality Assurance flashcards as text
  1. Which of the following is a key indicator that a transducer element has failed during QA testing?

    Answer: A dark band or dropout artifact in the image

    Failed transducer elements appear as dark vertical bands or dropout regions in the B-mode image because those elements no longer transmit or receive ultrasound.

  2. What does the term 'dynamic range' refer to in echocardiography system performance testing?

    Answer: The ratio of the largest to smallest signal amplitudes the system can display

    Dynamic range describes the span of echo signal amplitudes—from the weakest detectable to the strongest—that the system can display simultaneously without saturation or loss.

  3. A QA protocol requires checking 'slice thickness' (elevation resolution). Which phantom feature is used for this test?

    Answer: Anechoic spherical targets at varying depths

    Anechoic spherical cysts of known diameter at varying depths are used to assess slice thickness, because a thick beam slice causes fill-in artifact in small cysts.

  4. Which document should a sonographer consult to understand acceptable QA performance tolerances for a specific echocardiography unit?

    Answer: The equipment manufacturer's service manual and QA specifications

    The manufacturer's service manual and QA specifications define acceptable performance tolerances unique to each system model.

  5. What is the primary purpose of recording baseline QA measurements when a new echocardiography system is first installed?

    Answer: To establish reference values for detecting future performance degradation

    Baseline measurements at installation provide reference values so that future QA tests can identify whether performance has degraded over time.

  6. During color Doppler QA, a technician observes that aliasing occurs at lower-than-expected velocities. The most likely cause is:

    Answer: PRF set too low, reducing the Nyquist limit

    When PRF is set too low, the Nyquist limit decreases and aliasing occurs at lower velocities than expected, indicating a system or operator setting problem.

  7. Which of the following best describes 'contrast detail' performance in echocardiography QA?

    Answer: The system's ability to distinguish low-contrast lesions from background tissue

    Contrast detail performance measures how well the system detects subtle differences in echo amplitude between a lesion and surrounding tissue.