Testing and Quality Assurance Flashcards

7 cards from real ECHOCARDIOGRAM practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Testing and Quality Assurance flashcards as text
  1. Which phantom is commonly used to evaluate spatial resolution in echocardiography systems?

    Answer: AIUM 100-mm test object

    The AIUM 100-mm test object is a standard phantom used to assess spatial resolution, dead zone, and range accuracy in ultrasound systems.

  2. What is the purpose of performing a 'dead zone' test on an echocardiography transducer?

    Answer: To measure the minimum depth at which targets can be resolved near the transducer face

    The dead zone test identifies the minimum depth at which structures can be detected, which is limited by the transmit pulse length near the transducer face.

  3. When performing quality control on a Doppler system, which parameter verifies the accuracy of velocity measurements?

    Answer: Doppler frequency accuracy using a flow phantom

    A flow phantom with known velocities is used to verify that the Doppler system accurately measures flow velocities within acceptable tolerances.

  4. A technician notices inconsistent brightness across the image field during a routine QA check. This most likely indicates a problem with:

    Answer: Time-gain compensation uniformity

    Non-uniform brightness across the image field typically results from improper time-gain compensation (TGC) settings or a malfunctioning TGC circuit.

  5. Which statement best describes the concept of 'sensitivity' in echocardiography QA testing?

    Answer: The ability to detect small targets at maximum imaging depth

    Sensitivity in QA testing refers to the system's ability to detect low-level echo signals from small or deep targets at the limits of its imaging range.

  6. During equipment QA, a technician finds that the system's calipers consistently measure a 10 cm distance as 10.5 cm. This error is classified as:

    Answer: Systematic distance measurement error

    A consistent offset in distance measurement represents a systematic error, often caused by an incorrect assumed speed of sound in the system calibration.

  7. How frequently should echocardiography equipment undergo comprehensive preventive maintenance according to best practice guidelines?

    Answer: Annually or per manufacturer recommendation

    Comprehensive preventive maintenance should be performed annually or according to manufacturer specifications to ensure optimal equipment performance and safety.