  1. The FDA's 2023 final rule on cybersecurity for medical devices (Section 524B of FD&C Act) requires manufacturers to submit a Software Bill of Materials (SBOM). The primary purpose of the SBOM is to:

    Answer: Enable identification of software components and their known vulnerabilities

    An SBOM provides a comprehensive inventory of software components, enabling manufacturers and users to rapidly identify which products are affected when new vulnerabilities in third-party components are disclosed.

  2. Under ISO 14971:2019, the risk management process requires that residual risks be evaluated against:

    Answer: Benefits of the intended use, weighed using an overall risk-benefit analysis

    ISO 14971:2019 requires that overall residual risk be weighed against the overall benefit of the device's intended use, and the risk-benefit analysis must conclude that benefits outweigh risks.

  3. Which regulatory pathway is most appropriate for a novel Class III medical software device with no predicate device?

    Answer: Premarket Approval (PMA)

    Class III devices that cannot claim substantial equivalence and do not qualify for De Novo must submit a Premarket Approval (PMA) application, the most stringent FDA premarket review pathway.

  4. SOC 2 Type II reports differ from SOC 2 Type I reports primarily because Type II reports:

    Answer: Assess the design AND operating effectiveness of controls over a period of time

    SOC 2 Type II evaluates both the design suitability and the operating effectiveness of controls over an observation period (typically 6–12 months), whereas Type I assesses only design at a single point in time.

  5. A medical device manufacturer must retain Device History Records (DHRs) under 21 CFR Part 820 for a minimum of:

    Answer: The expected life of the device or 2 years from the date of release, whichever is longer

    21 CFR 820.184 requires DHRs to be retained for a period equivalent to the design and expected life of the device, but in no case less than 2 years from the date of release for commercial distribution.

  6. In a threat modeling exercise for a networked infusion pump, the STRIDE model categorizes 'an attacker forging dosage commands by impersonating a clinician workstation' under which threat category?

    Answer: Spoofing

    Spoofing in the STRIDE model refers to an attacker illegitimately assuming the identity of another user, system, or device—exactly what occurs when commands are forged by impersonating a legitimate clinician workstation.

  7. The EU MDR requires manufacturers of Class IIb and Class III devices to perform clinical follow-up through a Post-Market Clinical Follow-Up (PMCF) plan. When must this plan be updated?

    Answer: Continuously, as part of the ongoing post-market surveillance system

    Under EU MDR Article 83 and Annex XIV, PMCF is an ongoing process integrated into the post-market surveillance system, not a one-time or event-triggered activity.