  1. When the Doppler angle correction cursor is positioned at 90 degrees, the calculated velocity will be:

    Answer: Infinite or mathematically undefined

    At 90 degrees, cos(90°) = 0, making the denominator of the Doppler equation zero, which produces a mathematically undefined (infinite) velocity value.

  2. Increasing the wall filter setting on a spectral Doppler display will:

    Answer: Remove low-frequency signals from the spectral display

    The wall filter (high-pass filter) eliminates low-frequency Doppler signals produced by slow-moving vessel walls and surrounding tissue, reducing clutter near the baseline.

  3. A 'tardus parvus' waveform observed distal to a stenosis is characterized by:

    Answer: Decreased PSV with prolonged systolic acceleration time

    Tardus parvus (Latin: slow-small) describes the dampened, delayed post-stenotic waveform with reduced peak systolic velocity and a slow, rounded systolic upstroke.

  4. To correct aliasing on a spectral Doppler display, which of the following would NOT be an effective solution?

    Answer: Decreasing the sample volume gate size

    Decreasing sample volume size changes the amount of vessel sampled but does not affect the PRF or the Nyquist limit, making it ineffective for correcting aliasing.

  5. In a standard spectral Doppler display, the vertical (y-axis) represents:

    Answer: Doppler frequency shift (displayed as velocity after angle correction)

    The y-axis in spectral Doppler displays the Doppler frequency shift, which is converted to velocity using angle correction, while the x-axis represents time.

  6. On a spectral Doppler display, flow moving AWAY from the transducer will appear:

    Answer: Below the baseline as a negative deflection

    By convention, flow toward the transducer produces a positive frequency shift and appears above the baseline, while flow away from the transducer produces a negative shift and appears below the baseline.

  7. Which of the following best describes a low-resistance Doppler waveform?

    Answer: Continuous forward flow throughout diastole with elevated end-diastolic velocity

    Low-resistance organs (kidney, brain, liver) require constant blood supply and maintain continuous forward flow throughout diastole, producing waveforms with elevated end-diastolic velocities.