What happens to the ultrasound beam at an interface where the angle of incidence equals the critical angle?
Answer: Refraction bends the beam to 90°
At the critical angle, refraction causes the transmitted beam to travel along the interface at 90°, and total internal reflection occurs.
Which Doppler mode provides the best sensitivity for detecting slow-flow states such as organ perfusion?
Answer: Power Doppler
Power Doppler displays the amplitude (power) of the Doppler signal and is more sensitive to slow flow than color or pulsed wave Doppler.
In B-mode imaging, increasing the time gain compensation (TGC) at depth primarily compensates for:
Answer: Attenuation of the ultrasound beam
TGC amplifies returning echoes from deeper structures to compensate for the progressive attenuation of the beam as it travels through tissue.
A resistive index (RI) value close to 1.0 in a renal artery waveform suggests:
Answer: Absent or reversed diastolic flow
An RI near 1.0 indicates minimal or absent end-diastolic flow, suggesting markedly elevated distal resistance.
Which statement best describes lateral resolution in ultrasound?
Answer: It is determined by beam width perpendicular to the beam axis
Lateral resolution depends on beam width; the narrower the beam at a given depth, the better the lateral resolution.
In M-mode echocardiography, what does the vertical axis represent?
Answer: Depth (distance from transducer)
In M-mode, the vertical axis shows depth from the transducer while the horizontal axis represents time.
Which material property most determines the acoustic impedance of a tissue?
Answer: Density and speed of sound
Acoustic impedance (Z) is the product of tissue density (ρ) and the speed of sound (c): Z = ρc.